Vertical autoclave sterilizer standard operating procedures
1. Vertical autoclave sterilizer the normal working conditions:.
A) environmental temperature + 5 ℃ ~ + 40 ℃
B) the relative humidity ≤ 80% RH
C) atmospheric pressure is 70 kPa to 106 kPa
D) work power to AC220V + 22 v, 50 Hz + 1 Hz
2. Vertical autoclave sterilizer method of use
2.1 water: add distilled water to the container, and to ensure that
the water at least 3-5 cm higher than the electric heat pipe.
Note: due to will go to a certain amount of water consumption in
the process of sterilization, so every time when using foot water
should be in accordance with the above requirement, suggest using
distilled water.
2.2 Stacked: items will be disinfected after wound into the
sterilization in barrel, in order to avoid items plug valve was
laid, should leave some space between and package, and steam
osmosis, ensure the quality of sterilization.
Items within the stacking height should not exceed disinfection
disinfection Jane height.
3.3 sealing: disinfection of tanks into the container, then put the
container hose inserted sterilization barrels of semicircle groove.
willContainer lid and container for the ears of the groove, and
wrench will 6 to the wing bolt by to tighten, make the lid and
container closed.
3.4 Vertical autoclave sterilizer heating operating procedures:
A) insert the plug into the power socket, small will bleed valve
handle to facilitate the discharge of the cold air.
B) open preheating switch, preheat indicator, to have a nasty
through purge valve steam purge valve should be under the small
handle, as the heat continues, pressure gauge pointer began to rise
and the steam pressure sterilizer.
C) the pressure control knob to adjust the sterilizing position
(maximum pressure of
0. 142 Mpa)
D) will be the timer knob clockwise spiral refers to more than
forecast the location of the heat preservation time about 10
minutes, then swing to forecast the location of the heat
preservation time.
E) closed preheating switch, sterilizer into the constant
temperature constant pressureState, during the heating indicator,
destroy the alternate display (as a sterilizer for automatic
constant temperature constant pressure), when the desired
sterilization time, heating power supply will be automatically
closed, lights out, and make a sound prompt, closing time switch,
sterilization end.
Vertical Autoclave Sterilizer Note:
1) water overheat protection, such as container ignore aquatic or
appear in the process of sterilization and water deficiency
phenomenon, the sterilizationMachine will be automatic overheat
protection, heating sterilizer will automatically cut off the power
supply at this time and make a sound tip: after waiting for
container with water, cooling, are plugged in and set up the
sterilization temperature and time.
2) over current protection: when current overload, power fuse will
automatically cut off the power, the fuse capacity of OA.
3) safety valve: when pressure exceeds 0165 Ma, automatic pressure,
exhaust.